BioMime by Meril
BioMime Aura Sirolimus Eluting Coronary Stent System
BioMime by Meril • Cobalt Chromium • Sirolimus Coating
Drug Eluting StentsMeril Life Sciences
Sirolimus eluting coronary stent system with 65μm strut thickness, biodegradable polymer coating, morphology mediated expansion, and advanced delivery system
Product overview
Sirolimus eluting coronary stent system with 65μm strut thickness, biodegradable polymer coating, morphology mediated expansion, and advanced delivery system
World's first tapered sirolimus-eluting stent (SES). Variable diameter from proximal to distal. CoCr platform with biodegradable polymer. For tapered coronary anatomy where single diameter stents cause mismatch.
Technical specifications
| stent material | Cobalt Chromium (L605) |
| strut thickness | 65μm (0.065mm or 0.0026 inch) |
| stent diameter | 2.00, 2.25, 2.50, 2.75, 3.00, 3.50, 4.00, 4.50 mm |
| stent length | 8, 13, 16, 19, 24, 29, 32, 37, 40, 44, 48 mm |
| mean recoil | ≤4.0% |
| mean foreshortening | ≤1.0% |
| drug | Sirolimus |
| drug dose | 1.25 μg/mm² |
| polymer | Biodegradable & Biocompatible (PLLA + PLGA) |
| delivery system | Rapid Exchange |
| nominal pressure | 9 atm (Refer IFU) |
| rated burst pressure | Rated Burst Pressure: 14/16 atm |
| balloon overhang | ≤0.5 mm |
| shaft outer diameter | Proximal: 2.13 F, Distal: 2.7 F |
| catheter length | 142 cm |
| guide catheter compatibility | 5F (Min 1 D: 0.056 inch / 1.42 mm) |
| tip entry profile | 0.014 inch (0.36) |
| max guidewire | 0.014 inch (0.36) |
Clinical scope
Anatomical focus
- • coronary arteries