Agile Healthcare
AGILE HEALTHCAREMeril Authorized
CallWhatsAppBrowse
Myra BMS Bare Metal Stent
TEST_EDITED_BRAND

Myra BMS Bare Metal Stent

Hybrid Cell Cobalt-Chromium Coronary Stent with Laser Bonded Tip Technology

Coronary StentsMeril Life Sciences

Hybrid cell design bare metal stent with closed cells at proximal and distal ends for ostial scaffolding and open cells at mid segment for flexible navigation. Features latest generation cobalt-chromium alloy L605 with laser bonded tip technology.

Product overview

Hybrid cell design bare metal stent with closed cells at proximal and distal ends for ostial scaffolding and open cells at mid segment for flexible navigation. Features latest generation cobalt-chromium alloy L605 with laser bonded tip technology.

Technical specifications

stent materialCobalt Chromium L605
strut thickness120 µm (0.120 mm)
stent diameters mm5, 6, 7, 8, 9, 10
stent lengths mm17, 27, 37, 47, 57
mean recoil<4.0%
mean foreshortening≤1.0%
delivery systemOver-the-wire (OTW) 0.035"
nominal pressure8 - 10 atm
rated burst pressure11 - 14 atm
balloon overhang1.5 mm
shaft outer diameterProximal-5 Fr
balloon materialPolyamide - Semi Compliant
radiopaque markers2-Platinum Iridium swaged radiopaque markers
usable catheter length80 cm, 135 cm
min sheath compatibility6Fr for diameter: 5.00-9.00 mm, 7Fr for diameter: 10.00 mm
max guidewire compatibility0.035" (0.89 mm)

Clinical scope

Procedures

  • Percutaneous Coronary Intervention
  • Coronary Angioplasty
  • Coronary Revascularization

Anatomical focus

  • Coronary Arteries
  • Native Coronary Vessels

Indications & use cases

Bare Metal StentHybrid Cell DesignOstial ScaffoldingFlexible Navigation120 µm StrutOver-The-Wire Delivery0.035 inch Guidewire CompatibleLarge Vessel Stenting